Angiography Injector

Angiography Injector

System dimension: HxWxD=1040MM x 770MM x 630MM
Weight:23.5Kg
Electrical requirements: a.c. 220V/110V 50/60Hz
Flow rate: 0.1 to 50 ml/sec. in:
0.1ml/sec. increments between 0.1ml/sec. and 9.9ml/sec.
1ml/sec. increments between 10ml/sec. and 50ml/sec.
Syringe capacity: 150ml
Volume: 0ml to max, syringe capacity in 1ml increments.
Pressure limits: programmable from 100psi to 1200psi in 10psi increments.
Injection delay time: 0 to 300 seconds
X-ray delay time: 0 to 300 seconds
Rise time: 0.1 to 10 seconds in 0.1 increments
Autofill load rate: programmable
Injection capabilities: 1 to 6 phases
Storage capabilities: 100 protocols
Test injection: programmable
Pause phase: 0 second to 255 seconds in 1 second increments.
Output: injector ready, exposure trigger, and injecting signals
Input: injection start signal